Aoip FAST-CAL HTM 2010 Basic Fast Industrial Calibration


Aoip FAST-CAL HTM 2010  Basic Fast Industrial Calibration Image

Are you interested in the product Aoip FAST-CAL HTM 2010 Basic Fast Industrial Calibration from manufacturer Aoip with code FAST-CAL HTM 2010? Contact us now and get offer. Imtek Engineering, the fastest and most reliable industrial equipment supplier in the world, will offer you the best offer!

Get Offer With E-Mail: info@im-tek.com
Get a offer for Aoip FAST-CAL HTM 2010 Basic Fast Industrial Calibration from our live support team now!

Fast-Cal HTM 2010 is a basic fast industrial calibration dry block for temperature thermometers from -35°C (Tamb 20°C) to 140°C. This model is ideal for the validation of washer disinfectors, steam sterilizers and autoclaves. In place of a removable insert, it has a fixed block with holes for a reference probe and the type of test sensor commonly used in validation applications. Key features:Calibration volume: Fixed block of 5 holes of diameter: 3x 4.5 mm / 1x 6.5 mm / 1x 8 mm, depth: 145 mmLightweight and ruggedEasy to use on site Temperature range -35°C (Tamb 20°C) to 140°C Stability ±0.02°C Accuracy - Basic ±0.2°C Accuracy – Complete (with semi standard PRT) ±0.15°C Heating time -30°C to 140°C: 15 min Cooling time 140°C to 0°C: 15 min Calibration volume Fixed 3x 4.5 mm / 1x 6.5 mm / 1x 8 mm Depth: 145 mm Process input 4-20 mA (for On Site model only) General specifications Size 228 x 248 x 143 mm Weight 6.6 kg Power supply 115 VAC or 230 VAC, 50/60 Hz 150 Watts Communication ports RS 232