Pressure Dew Point Meter ZHM-100/450 EA2-ZHM-100
Parameters
Data
Sensor PDP
ZHM100, ceramic, traceable 13 point calibration
PDP range
-100°Cdp do +20°Cdp, 0 - 3000 ppmv
Flow rate
1- 5 Nlt./min
Pressure range
450 barg, max.
Operating temperature
-40°C to +60°C
Accuracy
+/- 1°C from -59°Cdp to +20°Cdp; +2/°C from -100°Cdp to -60°Cdp
Resolution
0.1°C from -79°Cdp do +20°Cdp; +/-1 digit; 1°C from -100°Cdp to -80°Cdp; +1/digit
Voltage
12-28 VDC
Current
20 mA, max.
Supply Voltage influence
+/- 0,005% RH/V
Temperature coeficient
Temperature compensated across operating temerature range
Protection
IP66 (NEMA 4)
Response time
5 mins to T95 (dry - wet)
Cable connection
Hirschamnn GDS series (DIN 4350-C)
Contacts
PIN1: 4-2mA, PIN3: power, PIN2: free
Weight
150 g
Sensor's calibration
by exchange
Housing material
AISI 316L
Dimensions
L= 132mm x pr. 27mm
Sensor filter
HDPE guard <10µm, optional: sintered guard 80µm