CALYS 100 is a field precision documenting multifunction calibrator within
CALYS range. It is the perfect tool for advanced process maintenance and use on
test bench in all industries. Suitable for all field and lab measurements, it
can simultaneously measure and generate over two isolated channels various
signals of temperature, resistance, process, pressure and frequency in one
single instrument.
Providing extended functionalities (temperature
simulation, scaling, steps, synthesizer, statistical functions…), CALYS 100
makes advanced data exploitation and full data traceability easier, as well as
quick access to functions by menus.
The instrument simultaneously measures and simulates:
·
Temperature: Up to 0,012% RDG
·
Resistance: Up to 0,010% RDG, 4 K? range
·
Current: Up to 0,012% RDG, 50 mA range +
24 V loop supply
·
Voltage: Up to 0,010% RDG, 50 V range
·
Frequency: Up to 0,005% RDG, 20 KHz
range (10 KHz in simulation)
·
Pressure: When used with an external
pressure module (ref. ACL433), CALYS 100 can measure and simulate pressure
(comparison calibration with a pressure pump).
Using this
user-friendly instrument, calibration tasks can be quickly carried out over the
whole process chain. Take the 900 g documenting process calibrator to the field
with you during the whole week with 10 calibration procedures stored in
the device. Run the procedure after connecting the probes to the instrument
(Easy connect system®) and save the results for onsite easy and quick
calibration.
Back to the office, you can then upload the data on a computer in order
to issue customized calibration certificates with dedicated
calibration software DATACAL.
IP 54, fully protected by an antichoc rubber holster, CALYS 100 integrates
"easyconnect" terminals and a wide backlite display that makes it
easy to use in any severe or dark conditions.
Easy connection system®
Connect your probes by simply pushing on the terminal top and insert wires
of up to 3 mm or 10 AWG diameter and compensated thermocouple connectors.
Wires are held tight between two brass plates ensuring thermal stability
and a very good cold junction compensation for thermocouples. This system also
enables 4 mm banana plugs and security connectors to be connected on the
terminal top.
CALYS series, 4 models from basic use to advanced performances
Specifications
CALYS 50
CALYS 75
CALYS 100
CALYS 150
Top accuracy
200 ppm
130 ppm
50 ppm
Temperature accuracy
Thermocouples (14) RTDs (12)
0.013% RDG for Tc K 0.012% RDG
0.01% RDG for Tc K 0.01% RDG
0.005% RDG for Tc K 0.006% RDG
DC current + Loop supply 24 V
Range Accuracy
50 mA 0.0175% RDG
100 mA 0.007% RDG
DC voltage
Range Accuracy
50 V IN / 20 V OUT 0.013% RDG
50 V 0.013% RDG
50 V 0.010% RDG
50 V 0.005% RDG
Frequency
Range Accuracy
20 KHz IN / 10 KHz OUT 0.005% RDG
100 KHz 0.01% RDG
Resistance
Range Accuracy
4000 ? 0.012% RDG
4000 ? 0.010% RDG
50 K? 0.006% RDG
Pressure
Range Accuracy
Relative pressure: 30 bar / Absolute pressure: 1,000 bar 0.05% RDG
Compliance to standards
21 CFR Part 11
NADCAP Heat treatment AMS 2750
Additional functions
Advanced data exploitation: Scaling, relative measurement, simulation of
ramps and steps, synthetizer, square root, statistical functions Transmitter
function
Additional functions
Switch test Calibration of transmitters
Additional functions
Comparison calibration HART: Digital calibration and data transfer
Calibration of thermistors
Software
DATACAL calibration software for configuration and data management
Memory
10,000 data stored and recalled on screen as curve or list