Kirchgaesser
Kirchgaesser MID-EX-E with CANopen Electromagnetic Small Flow Meter
Product Code:MID-EX-E with CANopen
Description
Measuring principle:Electromagnetic (flow)Thinfilm sensor (pressure)Thermistor (temperature)Flow measuring ranges:0...10 l/min to 0...600 l/minPressure measuring ranges:0...40 bar (0...4 MPa) to 0...400 bar (0...40 MPa)Temperature measuring ranges:0...100 °CNominal widths:DN10, DN19 and DN25Optional with plug couplers DN10, DN12, DN19, DN25 or DN31Optional with socket nipples DN32 and DN50 (type SSKV heavy version)Nominal pressure ranges:100 bar (10 MPa) to 500 bar (50 MPa)Electrical connection:Circular connector M12x1 (Binder series 763)Output signal:CANopenHousing:Sensor: brass or bronzeElectronic: plastic or brassElectrode material:1.4571 (316Ti)Power supply:MID-EX-EM*: 8.0...9.0 VDCMID-EX-ET*: 8.0...13.5 VDCPower consumption:< 1 WProtection acc. to EN 60529:IP 65Vibration resistance –
Endurance stress with sliding
frequency acc. to EN 60068-2-6:Excitation mode: sineFrequency range: 5...500 HzAmplitude: 5…15 Hz: 5.5 mm (peak), 15…500 Hz: 5 gdf/dt: 1 oct / minTest directions: 3 directions (X, Y, Z)Test duration: approx. 70 min / directionTotal test duration: approx. 3.5 hTest temperature: room temperatureShock resistance -
Acc. to EN 60068-2-27:Excitation mode: shock half – sineShock duration: 18 msShock amplitude: 30 gNo. of shocks: 3 per directionTest directions: 6 directions in space (± X, ± Y, ± Z)Total no. of shocks: 18Test temperature: room temperatureMarking acc. to 2014/34/EU (ATEX):I M2 Ex ia I Mb (BVS 07 ATEX E 060)Marking acc. to IECEx:Ex ia I Mb (IECEx BVS 07.0010, Dekra EXAM, Germany)Ex ia I (IECEx TSA 11.0043X, TestSafe, WorkCover NSW, Australia)Ambient temperature:0...80 °CProcess temperature:0...100 °CMeasuring uncertainty:Max. ± 2 % of end value, typ. ± 1 % of end value
Endurance stress with sliding
frequency acc. to EN 60068-2-6:Excitation mode: sineFrequency range: 5...500 HzAmplitude: 5…15 Hz: 5.5 mm (peak), 15…500 Hz: 5 gdf/dt: 1 oct / minTest directions: 3 directions (X, Y, Z)Test duration: approx. 70 min / directionTotal test duration: approx. 3.5 hTest temperature: room temperatureShock resistance -
Acc. to EN 60068-2-27:Excitation mode: shock half – sineShock duration: 18 msShock amplitude: 30 gNo. of shocks: 3 per directionTest directions: 6 directions in space (± X, ± Y, ± Z)Total no. of shocks: 18Test temperature: room temperatureMarking acc. to 2014/34/EU (ATEX):I M2 Ex ia I Mb (BVS 07 ATEX E 060)Marking acc. to IECEx:Ex ia I Mb (IECEx BVS 07.0010, Dekra EXAM, Germany)Ex ia I (IECEx TSA 11.0043X, TestSafe, WorkCover NSW, Australia)Ambient temperature:0...80 °CProcess temperature:0...100 °CMeasuring uncertainty:Max. ± 2 % of end value, typ. ± 1 % of end value
Technical Specifications
| Measuring principle: |
|
| Flow measuring ranges: | 0...10 l/min to 0...600 l/min |
| Pressure measuring ranges: | 0...40 bar (0...4 MPa) to 0...400 bar (0...40 MPa) |
| Temperature measuring ranges: | 0...100 °C |
| Nominal widths: |
|
| Nominal pressure ranges: | 100 bar (10 MPa) to 500 bar (50 MPa) |
| Electrical connection: | Circular connector M12x1 (Binder series 763) |
| Output signal: | CANopen |
| Housing: |
|
| Electrode material: | 1.4571 (316Ti) |
| Power supply: |
|
| Power consumption: | < 1 W |
| Protection acc. to EN 60529: | IP 65 |
| Vibration resistance – Endurance stress with sliding frequency acc. to EN 60068-2-6: |
|
| Shock resistance - Acc. to EN 60068-2-27: |
|
| Marking acc. to 2014/34/EU (ATEX): | I M2 Ex ia I Mb (BVS 07 ATEX E 060) |
| Marking acc. to IECEx: |
|
| Ambient temperature: | 0...80 °C |
| Process temperature: | 0...100 °C |
| Measuring uncertainty: | Max. ± 2 % of end value, typ. ± 1 % of end value |