Aoip FAST-CAL HIGH Basic Fast Industrial Calibration
Product Code:FAST-CAL HIGH
Description
Fast-Cal High is a basic fast industrial calibration dry block for temperature thermometers from 35°C to 650°C.
Fast-Cal Medium is available in three models, the BASIC (B), the SITE (S) and the SITE (S-TC).
The B model includes a sophisticated temperature controller with a dual display for set temperature and dry block temperature.
The S and S-TC models additionally include an in-built temperature digital indicator to which an external temperature sensor can be connected, giving greater accuracy eliminating temperature gradient and loading errors. The S model will accept a semi standard PRT, the S-TC model will work with a type N thermocouple.
Key features:Calibration volume: 25 x 148 mm depthLightweight and ruggedEasy to use on site Temperature range35°C to +650°CStability±0.03°C to ±0.05°CAccuracy - Basic±1°C at 500°CAccuracy – Complete (with semi standard PRT)±2°C at 650°CHeating time50°C to 650°C: 20 minCooling time650°C to 300°C: 20 minCalibration volume25 x 148 mm depthProcess input4-20 mA (for On Site model only)General specificationsSize228 x 248 x 143 mmWeight6.35 kgPower supply115 VAC or 230 VAC, 50/60 Hz750 WattsCommunication portsRS 232
Technical Specifications
Fast-Cal High is a basic fast industrial calibration dry block for temperature thermometers from 35°C to 650°C.
Fast-Cal Medium is available in three models, the BASIC (B), the SITE (S) and the SITE (S-TC).
The B model includes a sophisticated temperature controller with a dual display for set temperature and dry block temperature.
The S and S-TC models additionally include an in-built temperature digital indicator to which an external temperature sensor can be connected, giving greater accuracy eliminating temperature gradient and loading errors. The S model will accept a semi standard PRT, the S-TC model will work with a type N thermocouple.
Key features:
- Calibration volume: 25 x 148 mm depth
- Lightweight and rugged
- Easy to use on site
| Temperature range | 35°C to +650°C |
| Stability | ±0.03°C to ±0.05°C |
| Accuracy - Basic | ±1°C at 500°C |
| Accuracy – Complete (with semi standard PRT) | ±2°C at 650°C |
| Heating time | 50°C to 650°C: 20 min |
| Cooling time | 650°C to 300°C: 20 min |
| Calibration volume | 25 x 148 mm depth |
| Process input | 4-20 mA (for On Site model only) |
General specifications
| Size | 228 x 248 x 143 mm |
| Weight | 6.35 kg |
| Power supply | 115 VAC or 230 VAC, 50/60 Hz 750 Watts |
| Communication ports | RS 232 |