Greisinger GGA381 Atmospheric Oxygen Sensor
Greisinger
Product Code: GGA381
Artikelnummer: 610032
Brand: Greisinger
GET OFFER: info@imtek.com.tr
Description for Greisinger GGA381 Atmospheric Oxygen Sensor
Atmospheric oxygen sensors, closed sensor type with pressure connectionFor low oxygen concentrations, with pressure connection, suitable for GMH 369xGENERAL:For devices with external pressure port (GMH 5695/3695) this housing is optimal. Especially for systems with high or low pressure or with existing back pressure by flow.APPLICATION:It can be screwed airtight (Attention: Observe permissible operating pressure!). The device-pressure port is connected to the sensor pressure port. The device measures and compensates for the actual pressure at the sensor.SPECIFICATIONS:Sensor element: GOEL 381Oxygen-partial pressure probe, mounted in external sensor housing replaceable (temperature sensor mounted in housing)Specific features: for the lowest O 2 concentrations, for protective gases, in general, precise and very small measurements and above 35 vol .% O 2Measuring range:Partial oxygen pressure: 0 ... 1100 hPa O 2Oxygen concentration: 0.0 ... 100.0% O 2Response time: T 90 <10 s Accuracy (at 25 ° C, 1013 hPa) <2 % O 2 ± 0.1% O 2<25% O 2 ± 0.5% O 2Scope of supply: sensor, flow diverter, T-pieceRecommended Products For Greisinger