atmospheric oxygen sensor, closed sensor type with pressure connection
Universal application, diving gas, longlife, with pressure connection, suitable for GMH 569x
GENERAL:
For devices with external pressure port (GMH 5695/3695) this housing is optimal. Especially for systems with high or low pressure or with existing back pressure by flow.
APPLICATION:
It can be screwed airtight (Attention: Observe permissible operating pressure!). The device-pressure port is connected to the sensor pressure port. The device measures and compensates for the actual pressure at the sensor.
SPECIFICATIONS:
Sensor element:
GOEL 370
Oxygen-partial pressure probe, mounted in external sensor housing replaceable (temperature sensor mounted in housing)
Specific features:
Long service life, for protective gases with a high O 2 concentration and oxygen content <35 vol.% O 2
Measuring range:
Partial oxygen pressure:
0 ... 1100 hPa O 2
Oxygen concentration:
0.0 ... 100.0% O 2
Response time: T 90
<10 s
Accuracy (at 25 ° C, 1013 hPa)
<2% O 2 ± 0.2% O 2
<25% O 2 ± 0.5% O 2
> 25% O 2 ± 0.5% O2
Scope of supply:
sensor, flow diverter, T-piece