Measuring standard
DIADUR phase grating on glass
Signal periods
180 000
Accuracy of graduation
±0.9”
Interpolation error per signal period1)
±0.1”
Reference marks
One
Hub inside diameter
51.2 mm
Mech. permissible speed
? 1000 rpm
Moment of inertia
1.2 · 10-3 kgm2
Permissible axial motion of measured shaft
? ±0.05 mm
Interface
1 VPP
Cutoff frequency –3 dB
–6 dB
? 800 kHz
? 1.3 MHz
Electrical connection
With housing: Cable 1 m, with M23 coupling
Without housing: Via PCB connector, 12-pin (adapter cable ID 372164-xx)
Cable length
? 150 m (with HEIDENHAIN cable)
Voltage supply
DC 5 V ±0.5 V
Current consumption
? 250 mA (without load)
Vibration 55 to 2000 Hz
Shock 6 ms
? 50 m/s2 (EN 60 068-2-6)
? 1000 m/s2 (EN 60 068-2-27)
Operating temperature
0 °C to 50 °C
Protection* EN 60 529
Without housing: IP 00
With housing: IP 40
With housing and shaft seal: IP 64
Starting torque
–
0.25 Nm
Mass
3.0 kg
3.1 kg incl. housing
* Please select when ordering
1) Interpolation error within one signal period and the accuracy of the graduation together result in the encoder-specific error; for additional error from mounting and bearing of the measured shaft, see Measuring accuracy