Ricambi industriali e soluzioni ingegneristiche dal 2001.
Buhler GAS 222.20 Sample Gas Probe
Schede Tecniche / Documentazione Prodotto
Contattaci! 24 ore lavorative (esclusi weekend e festivi) per confermare disponibilità e tempi di consegna.
Descrizione del prodotto
I prodotti Bühler, incluso Buhler GAS 222.20 Sample Gas Probe, vengono approvvigionati tramite canali di importazione parallela e consegnati ai nostri clienti in United States con rapidità e affidabilità. Imtek Engineering offre componenti originali Bühler a prezzi competitivi. Anche se Buhler GAS 222.20 Sample Gas Probe non è elencato sul nostro sito, è possibile richiedere un preventivo in qualsiasi momento — il nostro team risponderà tempestivamente.
- Heated, temperature adjustable up to 200 °C / 392 °F
- 115/230 V
- Base unit
- With weather hood
- Heater with electronic regulator
- Limit outputs
- Insulation
- Including flange seal and screws
Sample gas probe with weather protection, temperature adjustable heater, and integrated particle filter. For applications with low dust load. The filter can be replaced without tools.
Technical Data Gas Probe Technical Data Probe operating temperature:
max. 200 °C
Ambient temperature without accessories:
-20 to +70 °C
Ambient temperature with accessories:
Component
Ambient temperature range
Compressed air valve:
-10 °C < Tamb < +55 °C
Pneumatic drive:
-20 °C < Tamb < +80 °C
Limit switch:
-20 °C < Tamb < +100 °C
Solenoid valve for pneumatic drive:
-10 °C < Tamb < +55 °C
Component
Medium temperature range
Compressed air valve:
-10 °C to +80 °C
Solenoid valve for pneumatic drive:
-10 °C to +100 °C
Medium temperature (blowback):
Regulator setting range:
+50 to +200 °C
Low/high temperature alarm:
Alarm adjustable ±5.....30 K from setpoint, factory preset to 15 K, max. switching
current 1 A
Electrical data:
230 V, 2.0 A, 50/60 Hz 115 V, 3.8 A, 50/60 Hz
Protection class:
IP54
Max. operating pressure:
6 bar
Materials in contact with media Flange:
Probe body:
Ball valve:
Seal:
Stainless steel 1.4571 Stainless steel 1.4571 Stainless steel 1.4408/1.4462/PTFE
Stainless steel 1.4404/graphite/and see filter Optional coating with SilcoNert® 2000 *
*The SilcoNert® 2000 coating forms a layer of hydrogenated, amorphous silicon on the surfaces of the measuring gas probe that
are in contact with media. It is applied by chemical vapour deposition (CVD), which creates a uniformly thin layer on the surface.
SilcoNert® 2000 is temperature-resistant up to 400 °C, chemically inert and hydrophobic. The coating is resistant to many chemically corrosive liquids. If you are unsure about the suitability for specific gas matrices, please contact Bühler Technologies
GmbH.
2
Bühler Technologies GmbH
We reserve the right to amend specification.
DE461021 ◦ 01/2026
Domande Frequenti
Domande frequenti su Buhler GAS 222.20 Sample Gas Probe di Bühler. Imtek Engineering è pronta ad assisterla nell'approvvigionamento di Buhler GAS 222.20 Sample Gas Probe da Bühler in United States.
Imtek Engineering si approvvigiona di prodotti originali Bühler tramite canali di fornitura affidabili e li offre ai clienti in United States a prezzi competitivi.