Industrial spare parts & engineering solutions since 2001.
Buhler GAS 222.11 Sample Gas Probe
기술 데이터시트 / 제품 문서
저희에게 연락하세요! 영업일 기준 24시간 이내(주말 및 공휴일 제외) 재고 상태와 예상 납기를 안내해 드립니다.
제품 설명
Bühler의 Buhler GAS 222.11 Sample Gas Probe을(를) 포함한 제품은 병행 수입 채널을 통해 조달되며 United States의 고객에게 신속하고 안정적으로 배송됩니다. Imtek Engineering은 경쟁력 있는 가격으로 정품 Bühler 부품을 제공합니다. Buhler GAS 222.11 Sample Gas Probe이 당사 웹사이트에 등록되어 있지 않더라도 언제든지 견적을 요청하실 수 있으며, 담당 팀이 신속히 답변 드릴 것입니다.
- Unheated
- Base unit with manual shut-off valve
- For in-situ and outlet filter
- With blowback connection
- The outlet filter can easily be removed by turning the handle 90°.
- For dust loads up to 2g/m³, non-condensable gases. Combined with inlet filter up to 10g/m³ and higher
- Including flange seal and screws
The probe has no innate ignition source and is therefore suitable for use in Ex areas.
The base unit becomes functional by adding accessories suitable for the application.
These include:
- Sampling tubes
- Unheated extensions
- Inlet filter
- Outlet filter
- Calibrating gas connection
- ANSI adapter flange
- Pneumatic drive
- Shut-off valve
- Blowback
- Blowback control
Technical Data Gas Probe Technical Data Probe operating temperature:
max. 200 °C
Ambient temperature without accessories:
-20 to +70 °C
Ambient temperature with accessories:
Component
Ambient temperature range
Compressed air valve:
-10 °C < Tamb < +55 °C
Pneumatic drive:
-20 °C < Tamb < +80 °C
Limit switch:
-20 °C < Tamb < +100 °C
Solenoid valve for pneumatic drive:
-10 °C < Tamb < +55 °C
Component
Medium temperature range
Compressed air valve:
-10 °C to +80 °C
Solenoid valve for pneumatic drive:
-10 °C to +100 °C
Medium temperature (blowback):
Regulator setting range:
+50 to +200 °C
Low/high temperature alarm:
Alarm adjustable ±5.....30 K from setpoint, factory preset to 15 K, max. switching
current 1 A
Electrical data:
230 V, 2.0 A, 50/60 Hz 115 V, 3.8 A, 50/60 Hz
Protection class:
IP54
Max. operating pressure:
6 bar
Materials in contact with media Flange:
Probe body:
Ball valve:
Seal:
Stainless steel 1.4571 Stainless steel 1.4571 Stainless steel 1.4408/1.4462/PTFE
Stainless steel 1.4404/graphite/and see filter Optional coating with SilcoNert® 2000 *
*The SilcoNert® 2000 coating forms a layer of hydrogenated, amorphous silicon on the surfaces of the measuring gas probe that
are in contact with media. It is applied by chemical vapour deposition (CVD), which creates a uniformly thin layer on the surface.
SilcoNert® 2000 is temperature-resistant up to 400 °C, chemically inert and hydrophobic. The coating is resistant to many chemically corrosive liquids. If you are unsure about the suitability for specific gas matrices, please contact Bühler Technologies
GmbH.
2
Bühler Technologies GmbH
We reserve the right to amend specification.
DE461021 ◦ 01/2026
자주 묻는 질문
Bühler의 Buhler GAS 222.11 Sample Gas Probe에 관한 자주 묻는 질문입니다. Imtek Engineering은 United States에서 Bühler의 Buhler GAS 222.11 Sample Gas Probe를 조달하는 데 도움을 드릴 준비가 되어 있습니다.
Imtek Engineering은 신뢰할 수 있는 공급 채널을 통해 정품 Bühler 제품을 조달하여 United States 고객에게 경쟁력 있는 가격으로 제공합니다.